Unitron TrueFit software instruction guide v2.2
35
To configure the function of the lever / dial on the hearing instrument(s) (if available):
1. Click the drop-down menu under the lever / dial label on the primary screen.
Choose from:
• Volume Control: the lever / dial adjusts the volume level of the hearing
instruments.
Note: DuoLink is disabled by default to allow independent adjustment of this
control on each ear.
• SmartFocus: the lever / dial adjusts the SmartFocus setting in the hearing
instrument.
Note: DuoLink is disabled by default to allow independent adjustment of this
control on each ear.
• Off: the lever / dial is disabled on the hearing instrument.
Note: The + / – buttons and scroll wheel on the Smart Control remote are not
impacted by the setting of this control.
Note: Only one user control on the hearing instrument can be assigned to perform
a particular function. While both the push button and lever/dial can be configured
as either a volume control or a SmartFocus control, both controls cannot perform
the same function.
To configure the function of the Micro CIC remote (if available):
1. Click the drop-down menu under the Micro CIC remote function label on the
primary screen. Choose from:
• Off: the micro CIC remote is disabled.
• Program change: bring the remote up to each hearing instrument to toggle
through the various listening programs.
• Volume up: bring the remote up to each hearing instrument to increase the
volume in steps. Once the hearing instrument reaches the maximum volume,
the next step will bring it back to the default volume level.
• Volume down: bring the remote up to each hearing instrument to decrease the
volume in steps. Once the hearing instrument reaches the minimum volume,
the next step will bring it back to the default volume level.
Volume Settings
1. Click the Volume Settings task item at the bottom of the screen.
2. Select the desired amount of increase in volume using the Volume Control Up
drop-down menu.
3. Select the desired amount of decrease in volume using the Volume Control Down
drop-down menu.
Note: The value selected in the volume control up and down boxes determines the
volume control range provided to the client (as displayed in the volume settings task
box). Each adjustment of the volume control using the Smart Control remote or on the
hearing instruments is 2 dB.
Start up Settings
1. Click the Start up Settings task item at the bottom of the screen.
2. Select the desired Start up Program from the drop-down menu.
3. Select the desired Start up Delay from the drop-down menu. Choose Minimum,
10 second delay or 15 second delay.