Unitron TrueFit software instruction guide v2.2
37
Beep Setup
1. Click the Beep Setup task item at the bottom of the screen.
2. Select the intensity level from the drop-down menu. Choose from:
Very Soft, Soft, Moderate, Loud, or Very Loud.
3. Select the frequency from the drop-down menu. Choose from: Low,
Medium or High.
Beep Enabling
1. Click the Beep Enabling task item at the bottom of the screen.
2. Select the beep notification types you would like active for the client.
Choose from: Program Change, Start Up Sound (initial sound which
occurs upon the powering up of the hearing instruments), Easy /
Automatic Events (events listed under the demo drop-down menu for
this category), VC & SmartFocus Min/Max (end of range), VC &
SmartFocus steps (beeps that occur each time these user controls are
changed), and Warning / Alerts (events listed under the demo drop-
down menu for this category).
Note: The beeps that play at the end of the trial period with Flex hearing
instruments cannot be disabled. When the end of trial period beeps start
to sound, fitters can turn off the repeated beeps by resetting the trial
period, i.e. by detecting and then saving to the hearing instruments.
Trial Duration
Note: This option is available with Flex devices.
1. Click the Trial Duration task item at the bottom of the screen.
2. Select the desired trial duration before the 'End of trial period' beeps
are generated in the hearing instruments.
Note: Once the trial duration has been exceeded, repeated beeps (End of
trial period beeps) will be generated. Initially, these beeps will occur every
hour for ~ 3 days of use, after which the instruments will repeatedly
generate the beeps every minute.
Fitting Summary
The Fitting Summary screen provides a succinct
overview of primary fitting-related items in an
easy-to-read table for each hearing instrument.
The table provides an opportunity for a final
check of the fitting settings and general status
prior to printing, saving and closing the session.
The default table information is dependent on the
product being fitted and can include: Program
List, Client Controls, General Status, Accessories,
and Indicators. If desired, additional details can
be obtained by clicking the More Info button
which then displays Fitting Settings and
Acoustics. Items outside the primary viewing area
can be observed by adjusting the vertical
scrolling slider.